Data Foundry

CPSC recalls 2012

TRUMPF Medical Systems Recalls Television Mounting System Due to Falling Hazard

U.S. Consumer Product Safety Commission recall (2012-07-18). Hazards: fall. About 215 units in the US.

AgencyU.S. Consumer Product Safety Commission
Recall number12742
Published2012-07-18
Last updated2016-04-21
Product categoryTelevisions, Mounts, Guards & Antennas
ProductsOndal AC2000 (PDI) Television Mounting System
Firm named in the titleTRUMPF Medical Systems
FirmsOndal Medical Systems of America Inc., of Sandston, VA (importer); TRUMPF Medical Systems Inc., of Charleston, SC (importer); TRUMPF Medical Systems Inc., of Charleston, SC (distributor)
Manufactured inGermany
Units (US)215
Hazard classesfall
Remedy classesrepair
SoldTRUMPF Medical Systems Inc. nationwide From January 2005 through December 2011 for about $2,100.

Description

The mounting systems were sold for use in hospital Intensive Care Units. The units are off-white and have multiple joints between the ceiling mount and the TV mount at the end of the arm system.

Hazard

The arm joints on the television mounting system do not have a washer at the top joint which can result in excessive wear of the stop pins on the arm. This can allow the arm system and items connected to the arm to fall and injure the user or bystanders.

Remedy

Secure the unit to prevent movement and contact TRUMPF Medical Systems for information on receiving a repair kit that includes the missing washers.

Incidents and injuries

The firm received one report of a mounting system falling while being positioned. No injuries have been reported.