Drive DeVilbiss Healthcare Recalls Adult Portable Bed Rails After Two Deaths; Entrapment and Asphyxiation Hazards
U.S. Consumer Product Safety Commission recall (2021-12-06). Hazards: entrapment, suffocation. Models: RTL15063-ADJ, RTL15073, 15062. About 496,100 units in the US and 68,000 in Canada.
| Agency | U.S. Consumer Product Safety Commission |
|---|---|
| Recall number | 22025 |
| Published | 2021-12-06 |
| Last updated | 2021-12-06 |
| Products | Bed Assist Handles and Bed Assist Rail |
| Firm named in the title | Drive DeVilbiss Healthcare |
| Firms | Drive DeVilbiss Healthcare, of Port Washington, New York (distributor) |
| Units (US) | 496,100 |
| Units (Canada) | 68,000 |
| Units (Mexico) | 119 |
| Model numbers | RTL15063-ADJ, RTL15073, 15062 |
| Hazard classes | entrapment, suffocation |
| Remedy classes | refund, stop use |
| Sold | Medical supply stores nationwide and online at www.amazon.com and www.walmart.com. The bed rails were sold from October 2007 through December 2021 for between $30 and $80. |
Description
This recall involves four models of Drive DeVilbiss Healthcare's adult portable bed rails: Model Product Product as Sold on Amazon: Dimensions 15064 Bed Assist Handle Home Bed Assist Handle About 21 inches high and 12 inches wide 15062 Bed Assist Rail with Folding Board Home Bed Assist Grab Rail with Bed Board About 23 inches high and 12 inches wide RTL15073 Home Bed Assist Handle Home Bed Assist Handle About 13-17 inches high (adjustable) and 19 inches wide RTL15063-ADJ Home Bed Assist Handle Adjustable Height Home Bed Assist Handle About 15-20 inches high (adjustable) and 20 inches wide The name "Drive" and the model number are printed on a label located on the product's metal tubing. The rails are made of steel tubing either in white or chrome. Models RTL15063-ADJ and RTL15073 include black non-slip padding on the grip handle and under-bed frame. Model 15062 features a wooden under-bed board attached to the grip handle.
Hazard
When attached to an adult's bed, users can become entrapped within the bed rail or between the rail and the side of the mattress. This poses a serious entrapment hazard and risk of death by asphyxiation.
Remedy
Consumers should immediately stop using the recalled bed rails and contact Drive DeVilbiss Healthcare to receive a full refund.
Incidents and injuries
Drive has received two reports of entrapment deaths associated with two of its Bed Assist Handle bed rails (models 15064 and RTL15073). The deaths occurred in February 2011 and February 2015, and involved a 93-year-old woman at her home in California, and a 92-year-old man at an assisted living facility in Canada.