Siemens AXIOM Luminos dRF Max systems The Axiom Luminos dRF is… — Class II medical device recall Z-0003-2015

Terminated recall by Siemens Medical Solutions USA, Inc, reported 2014-10-08, distributed nationwide. Two malfunctions posing potential risk to patients were identified with Siemens Ysio Max, AXIOM Luminos dRF Ma

Recall numberZ-0003-2015
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSiemens Medical Solutions USA, Inc, Malvern, PA, United States
Recall initiated2014-09-02
Classified2014-10-02
Reported by FDA2014-10-08
Terminated2015-02-24
DistributedNationwide (US)
Quantity1
Reason classeslabeling, device malfunction
Serial numbers5037

Product

Siemens AXIOM Luminos dRF Max systems The Axiom Luminos dRF is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic studies. Using a digital flat detector, it can perform a range of applications including general R/F, angiography and pediatric examinations. The Axiom Luminos dRF is a device intended to visualize anatomical structures by converting a pattern of X-ray into a visible image.

Reason for recall

Two malfunctions posing potential risk to patients were identified with Siemens Ysio Max, AXIOM Luminos dRF Max, Uroskop Omnia Max, and Luminos Agile Max systems with a specific set of serial numbers that will cause lost images due to an automatic Radiology Information System (RIS) worklist update, thus causing examinations to be repeated. Image label may also be displayed incorrectly after images