Miami J Select Collar — Class II medical device recall Z-0006-2025
Ongoing recall by Ossur H / F, reported 2024-10-09, distributed in AK, AL, AR, AZ, CA, CO, CT, FL…. Pressure sensitive adhesive hooks attached to the sides of the anterior cervical orthosis collar panel can occ
| Recall number | Z-0006-2025 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Ossur H / F, Reykjavik, N/A, Iceland |
| Recall initiated | 2024-08-20 |
| Classified | 2024-10-01 |
| Reported by FDA | 2024-10-09 |
| Distributed | AK, AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MT, NC, NE, NH, NJ, NM, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA |
| Quantity | 21,020 |
| UPC / GTIN / UDI-DI | 05690967817008, 05690967817145 |
| Model numbers | MJS-101, MJSR-101 |
Product
Miami J Select Collar, Catalog: MJS-101; Miami J Select Collar Set, Catalog: MJSR-101
Reason for recall
Pressure sensitive adhesive hooks attached to the sides of the anterior cervical orthosis collar panel can occasionally exhibit poor adhesion, which may lead to them gradually peeling off with repeated doffing, possibly leading to reduced immobilization of the cervical spine.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0006-2025. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.