Data Foundry

Medical device recalls 2014

VITROS Chemistry Products PHYT — Class II medical device recall Z-0007-2015

Terminated recall by Ortho-Clinical Diagnostics, reported 2014-10-08, distributed nationwide. Ortho Clinical Diagnostics identified a potential for biased results to be generated when using VITROS PHYT Sl

Recall numberZ-0007-2015
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmOrtho-Clinical Diagnostics, Rochester, NY, United States
Recall initiated2014-08-20
Classified2014-10-02
Reported by FDA2014-10-08
Terminated2017-04-11
DistributedNationwide (US)
Countries outside the USSG
Lot numbers2613-0150-4913
Expiration dates2014-11-15

Product

VITROS Chemistry Products PHYT (Phenytoin) Slides, REF/Product Code 829 8671, IVD. For in vitro diagnostic use only. VITROS Chemistry Products PHYT Slides quantitatively measure phenytoin (PHYT) concentration in serum and plasma using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.

Reason for recall

Ortho Clinical Diagnostics identified a potential for biased results to be generated when using VITROS PHYT Slides, Lot 2613-0150-4913. Some positively biased results were identified during testing of a different VITROS PHYT Slide lot; the affected slides were not released for distribution. Since Lot 2613-0150-4913 was manufactured in the same event, there is the potential for slides from this lot to also be affected.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0007-2015. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.