Data Foundry

Medical device recalls 2024

Tablo Console — Class II medical device recall Z-0019-2025

Ongoing recall by Outset Medical, Inc., reported 2024-10-09, distributed in AK, AL, AR, AZ, CA, CO, CT, DE…. Blood leak sensor, a component of hemodialysis device, may show decreased sensitivity, at the lowest level, sp

Recall numberZ-0019-2025
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmOutset Medical, Inc., San Jose, CA, United States
Recall initiated2024-08-01
Classified2024-10-03
Reported by FDA2024-10-09
DistributedAK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV, WY
Countries outside the USAE
Quantity2,422
Reason classespackaging
UPC / GTIN / UDI-DI00850001011112, 00850001011204
Serial numbers400048, 400058, 400063, 400082, 400087, 400109, 400141, 400205, 400222, 400227, 400288, 400317, 400347, 400394, 400405, 400412, 400424, 400425, 400432, 400440, 400463, 400487, 400849, 400916, 400981 and 475 more
Model numbersPN-0003000, PN-0006000

Product

Tablo Console, REF: PN-0003000, PN-0006000, part of the Tablo Hemodialysis System. The Tablo Hemodialysis System is indicated for use in patients with acute and/or chronic renal failure, with or without ultrafiltration, in an acute or chronic care facility.

Reason for recall

Blood leak sensor, a component of hemodialysis device, may show decreased sensitivity, at the lowest level, specified as 0.35 mL/min blood leak rate, which could delay triggering the ALARM_DIALYZER_BLOOD_LEAK alarm, if blood loss, less-than 300ml, occurs, the patient may experience hypotension, fatigue, cramps, dizziness, headaches, nausea, and shortness of breath.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0019-2025. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.