Data Foundry

Medical device recalls 2012

The Eon Mini Implantable Pulse Generator — Class II medical device recall Z-0025-2013

Terminated recall by Advanced Neuromodulation Systems Inc., reported 2012-10-17, distributed nationwide. Eon Mini IPGs manufactured in April 2012 could potentially exhibit a sudden, brief surge in stimulation that w

Recall numberZ-0025-2013
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmAdvanced Neuromodulation Systems Inc., Plano, TX, United States
Recall initiated2012-09-04
Classified2012-10-09
Reported by FDA2012-10-17
Terminated2014-05-12
DistributedNationwide (US)
Quantity56 units

Product

The Eon Mini Implantable Pulse Generator (IPG) is a 16-channel, rechargeable, multi-program system that allows the connection of one or two leads with four, eight, and 16 electrode leads. It is powered by a hermetically sealed battery within a titanium case. It uses microelectronic circuitry to generate constant current electrical stimulation. Stimulation programs can be delivered as either single stimulation or MultiStim programs depending on patients need. Product Usage: Eon Mini system is indicated as an aid in the management of chronic intractable pain of the trunk and/or limbs including unilateral or bilateral pain associated with the following: failed back surgery syndrome, intractable low back and leg pain.

Reason for recall

Eon Mini IPGs manufactured in April 2012 could potentially exhibit a sudden, brief surge in stimulation that would be felt by the patient as uncomfortable or painful. SJMs routine testing, monitoring, and analysis of manufacturing processes identified three lots of Eon Mini IPGs that the internal battery has the potential to come into contact with the internal microcontroller board.