Data Foundry

Medical device recalls 2013

Boston Scientific COGNIS CRT-D — Class II medical device recall Z-0025-2014

Terminated recall by Boston Scientific CRM Corp, reported 2013-10-23, distributed nationwide. The performance of a low voltage capacitor in a subset of COGNIS CRT-Ds and TELIGEN ICDs manufactured prior to

Recall numberZ-0025-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmBoston Scientific CRM Corp, Saint Paul, MN, United States
Recall initiated2013-08-29
Classified2013-10-17
Reported by FDA2013-10-23
Terminated2014-04-10
DistributedNationwide (US)
Countries outside the USAR, AT, AU, AW, BE, BM, BR, CA, CH, CL, CO, CR, CY, CZ, DE, DK, EC, EG, ES, FI, FR, GB, GR, HK, HU, IE, IL, IR, IT, JO, JP, KR, KW, LB, LV, MX, MY, NL, NO, NZ, PA, PE, PL, PT, SA, SE, SG, SI, SK, SV, TH, TT, TW, VE, ZA
Quantity38,500 worldwide
Serial numbers000036, 000085, 000087, 000088, 000090, 000091, 000093, 000094, 000096, 000097, 000098, 000103, 000106, 000107, 000112, 000114, 000115, 000117, 000119, 000120, 000121, 000125, 000128, 000132, 000133 and 475 more
Model numbersN106, N107, N118, N119, P106, P107

Product

Boston Scientific COGNIS CRT-D, Cardiac Resynchronization Therapy Defibrillator, models N118, N119. The following models are not available in the US: models N106, N107, P106, P107. Sterilized using ethylene oxide. Boston Scientific, 4100 Hamline Avenue North, St. Paul, MN 55112-5798 USA. Indicated for patients who are at risk for sudden cardiac death due to ventricular arrhythmias and may require pacing support.

Reason for recall

The performance of a low voltage capacitor in a subset of COGNIS CRT-Ds and TELIGEN ICDs manufactured prior to December 2009, may be compromised over time, causing increased current drain that can lead to premature battery depletion.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0025-2014. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.