Data Foundry

Medical device recalls 2024

BD Synapsys Informatics Solution with software versions: 4 — Class II medical device recall Z-0026-2025

Ongoing recall by Becton Dickinson & Co., reported 2024-10-16, distributed nationwide. Synapsys software versions 4.20, 4.30, 5.10, 5.20, or 5.30 and utilizing the in-line breakpoint feature, an er

Recall numberZ-0026-2025
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmBecton Dickinson & Co., Sparks, MD, United States
Recall initiated2024-09-12
Classified2024-10-08
Reported by FDA2024-10-16
DistributedNationwide (US)
Countries outside the USAT, AU, BE, CA, CH, CN, DE, DK, ES, FR, GB, HK, IT, JP, LU, MX, NL, NO, NZ, RU, SA, SE, SG, TW, ZA
Quantity162 units
Reason classessoftware
UPC / GTIN / UDI-DI00382904441500
Model numbers444150

Product

BD Synapsys Informatics Solution with software versions: 4.20, 4.30, 5.10, 5.20 and 5.30- IVD standalone laboratory software solution providing data management and workflow management functionality across locations where clinical diagnostic activities take place. Catalog Number: 444150

Reason for recall

Synapsys software versions 4.20, 4.30, 5.10, 5.20, or 5.30 and utilizing the in-line breakpoint feature, an error is occurring where BD Synapsys will incorrectly display breakpoints for organisms that do not have defined breakpoints in CLSI or EUCAST. This occurs when the identified organism is included in the same interpretive group as a species for which there is a specific breakpoint. May result in either a false susceptible or false resistant result. A false susceptible result can cause inadequate or incomplete antibiotic treatment due to the use of an ineffective antibiotic, and a false resistant result may limit access to a preferred antibiotic regimen with possible elongation of therapy

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0026-2025. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.