Data Foundry

Medical device recalls 2012

CYTO-STAT tetraCHROME CD45-FITC/CD4-RD1/CD8-ECD/CD3-PC5 — Class II medical device recall Z-0033-2013

Terminated recall by Beckman Coulter Inc., reported 2012-10-17, distributed nationwide. The recall was initiated because Beckman Coulter has determined that the specimen and prepared sample stabilit

Recall numberZ-0033-2013
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmBeckman Coulter Inc., Brea, CA, United States
Recall initiated2012-05-22
Classified2012-10-10
Reported by FDA2012-10-17
Terminated2012-10-10
DistributedNationwide (US)
Countries outside the USAE, AT, AU, BB, BD, BE, BG, BH, CA, CH, CL, CN, CO, CZ, DE, DK, EC, ES, FI, FR, GB, GR, HK, HR, ID, IE, IL, IN, IQ, IT, JP, KE, KR, KW, LT, MA, MX, MY, NG, NL, NO, NZ, OM, PA, PK, PL, QA, RO, SA, SE, SG, SI, SK, SV, TR, TW, UY, ZA
Reason classeslabeling, specification failure, software
Model numbers6607013

Product

CYTO-STAT tetraCHROME CD45-FITC/CD4-RD1/CD8-ECD/CD3-PC5, Part Number 6607013. Allows simultaneous identification and enumeration of total CD3+, total CD4+, total CD8+, dual CD3+/CD4+, dual CD3+/CD8+ and/or total CD3+, CD19+ and CD3+/CD56+ lymphocyte percentages and absolute counts in whole blood by flow cytometry. The systems also provide the CD4/CD8 ratio when using CD45-FITC/ CD4-RD1/CD8-ECD/CD3-PC5, and total lymphocyte percentage when using CD45-FITC/CD56-RD1/ CD19-ECD/CD3-PC5.

Reason for recall

The recall was initiated because Beckman Coulter has determined that the specimen and prepared sample stability claims stated in the product labeling for tetraCXP software and CYTO-STAT reagents (CD45-FITC/CD4-RD1/CD8-ECD/CD3-PC5 and CD45-FITC/CD56-RD1/CD19-ECD/CD3-PC5) need to be updated for alignment with current standards and clinical decision-making associated with lymphocyte immunophenotyping

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0033-2013. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.