Data Foundry

Medical device recalls 2013

Xoran Mini CT Scanner — Class II medical device recall Z-0036-2014

Terminated recall by Xoran Technologies, Inc., reported 2013-11-06, distributed nationwide. A description of the defect in the product or the manner in which the product fails to comply with an applicab

Recall numberZ-0036-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmXoran Technologies, Inc., Ann Arbor, MI, United States
Recall initiated2010-07-27
Classified2013-10-25
Reported by FDA2013-11-06
Terminated2014-09-11
DistributedNationwide (US)
Quantity320

Product

Xoran Mini CT Scanner

Reason for recall

A description of the defect in the product or the manner in which the product fails to comply with an applicable Federal standard; The MiniCAT CT Scanners subject to this notification failed to meet Xoran's specifications for dose rate of 125 ¿¿ 3 kVp. These 18 complaints received include dose rates both on the high and low end of Xoran's specifications. See ATTACHMENT 1 for details. (e) A n evalu

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0036-2014. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.