Meera Mobile Operating Table-to support and position the patient… — Class II medical device recall Z-0036-2024
Ongoing recall by Getinge Usa Sales Inc, reported 2023-10-11, distributed nationwide. Under certain conditions, an issue might prevent the device from performing as intended. A specific sequence o
| Recall number | Z-0036-2024 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Getinge Usa Sales Inc, Wayne, NJ, United States |
| Recall initiated | 2023-09-06 |
| Classified | 2023-10-04 |
| Reported by FDA | 2023-10-11 |
| Distributed | Nationwide (US) |
| Quantity | 224 units |
| UPC / GTIN / UDI-DI | 04046768124363 |
| Serial numbers | 100, 101, 102, 103, 104, 105, 106, 107, 108, 110, 111, 112, 113, 114, 115, 116, 118, 119, 123, 124, 125, 126, 127, 128, 129 and 124 more |
| Model numbers | 7200.01F2 |
Product
Meera Mobile Operating Table-to support and position the patient immediately before, during and after surgical interventions as well as for examination and treatment. Model Number: 7200.01F2
Reason for recall
Under certain conditions, an issue might prevent the device from performing as intended. A specific sequence of commands on the control unit activates the traction drive and triggers an unintended driving (autodrive) of max. 7 sec. may result in the following reasonably foreseeable injury or damage to the health of patients and/or users
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0036-2024. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.