Data Foundry

Medical device recalls 2025

This device is a digital radiography/fluoroscopy system used in a… — Class II medical device recall Z-0037-2026

Ongoing recall by Canon Medical System, USA, INC., reported 2025-10-15. It has been found that when SPOT Fluoro is first turned on, the default ROI is selected and the collimator bla

Recall numberZ-0037-2026
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedFDA mandated
Recalling firmCanon Medical System, USA, INC., Tustin, CA, United States
Recall initiated2025-09-02
Classified2025-10-08
Reported by FDA2025-10-15
Quantity8
Reason classessoftware
Serial numbersBBF2392003, SUC2392013, SUC23Z2014, SXA2432001, SXA2432002, SZA2422001, SZA2422002, SZA2442004

Product

This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities.

Reason for recall

It has been found that when SPOT Fluoro is first turned on, the default ROI is selected and the collimator blades move to the boundary of the ROI, and the static image is overlayed around the ROI, covering the collimated area. If the customer changes the ROI, the collimator blades and static images are adjusted accordingly. However, when the acquisition program is changed, the collimator returns to the default ROI position which may be larger than the specified one, leaving the X-ray area partially covered by the static image. This is a software problem with the SPOT Fluoro program and will not occur if the function is not used.