GE Healthcare Centricity Universal Viewer Zero Footprint Client — Class II medical device recall Z-0039-2025
Ongoing recall by GE Healthcare, reported 2024-10-16. GE HealthCare has become aware of an issue in Centricity Universal Viewer Zero Footprint Client (ZFP) version
| Recall number | Z-0039-2025 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | GE Healthcare, Chicago, IL, United States |
| Recall initiated | 2024-08-29 |
| Classified | 2024-10-09 |
| Reported by FDA | 2024-10-16 |
| Quantity | 1,500 units |
| UPC / GTIN / UDI-DI | 00840682102988 |
Product
GE Healthcare Centricity Universal Viewer Zero Footprint Client, Model/Catalog/Code Numbers: a)¿¿¿¿¿ 2089507-120; b)¿¿¿¿ 2089507-129; c)¿¿¿¿¿ 2089507-131; d)¿¿¿¿ 2089507-135; e)¿¿¿¿ 2089507-136; f)¿¿¿¿¿¿ 2089507-140; g)¿¿¿¿¿ 2089507-144; h)¿¿¿¿ 2089507-145; i)¿¿¿¿¿¿ 2089507-149; j)¿¿¿¿¿¿ 2089507-153; k)¿¿¿¿¿ 2089507-154; l)¿¿¿¿¿¿ 2089507-155; m)¿¿¿ 2089507-156; n)¿¿¿¿ 2089507-157; o)¿¿¿¿ K2037DAED; p)¿¿¿¿ K2037DBED; q)¿¿¿¿ K2037DED; r)¿¿¿¿¿ K2037DJED; s)¿¿¿¿¿ K2037DNED; t)¿¿¿¿¿ K2042TTED; u)¿¿¿¿ K2049DC; v)¿¿¿¿¿ K2049DD; w)¿¿¿ K2049DF; x)¿¿¿¿¿ P2037DAED; y)¿¿¿¿¿ P2037DBED; z)¿¿¿¿¿ P2037DED; aa)¿¿ To be provided
Reason for recall
GE HealthCare has become aware of an issue in Centricity Universal Viewer Zero Footprint Client (ZFP) version v6.0 SP11.x where the latest report is not shown by default to the user. The issue may occur when a study has multiple DICOM structured reports (i.e. preliminary, final, and/or addendum) stored in different series.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0039-2025. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.