Data Foundry

Medical device recalls 2013

Xoran MiniCAT00202 — Class II medical device recall Z-0040-2014

Terminated recall by Xoran Technologies, Inc., reported 2013-11-06, distributed nationwide. The MiniCAT CT Scanners subject to this notification failed to meet Xoran's specifications for dose rate of 12

Recall numberZ-0040-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmXoran Technologies, Inc., Ann Arbor, MI, United States
Recall initiated2010-07-27
Classified2013-10-25
Reported by FDA2013-11-06
Terminated2014-09-11
DistributedNationwide (US)
Quantity320

Product

Xoran MiniCAT00202

Reason for recall

The MiniCAT CT Scanners subject to this notification failed to meet Xoran's specifications for dose rate of 125 ¿ 3 kVp. These 18 complaints received include dose rates both on the high and low end of Xorans specifications.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0040-2014. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.