Alere Triage¿ TOX Drug Screen Control 1 Product Generic Name: Quality… — Class III medical device recall Z-0053-2014
Terminated recall by Alere San Diego, Inc., reported 2013-10-30, distributed nationwide. Alere San Diego, Inc. initiated a voluntary recall of Alere Triage TOX Drug Screen Control 1 (PN 94413, Lot C2
| Recall number | Z-0053-2014 |
|---|---|
| Product type | Medical device |
| Classification | Class III |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Alere San Diego, Inc., San Diego, CA, United States |
| Recall initiated | 2013-09-23 |
| Classified | 2013-10-23 |
| Reported by FDA | 2013-10-30 |
| Terminated | 2014-09-26 |
| Distributed | Nationwide (US) |
| Countries outside the US | CA, ZA |
| Quantity | 38 |
| Lot numbers | C2681A |
| Model numbers | 94413 |
Product
Alere Triage¿ TOX Drug Screen Control 1 Product Generic Name: Quality Control Material Description of the product: Good Laboratory Practice suggests that external controls should be tested with each new lot or shipment of test materials, or every 30 days, and as otherwise required by a laboratorys standard quality control procedures. The Alere Triage¿ Drug Screen Controls, when used in accordance with the test procedures described in the Alere Triage¿ TOX Drug Screen package insert, provide this capability. The results obtained for the controls are to be compared with the assigned values given on the Expected Values Card, accompanying the package insert, to determine if the procedure is within control limits. Alere Triage¿ TOX Drug Screen Control 1 or Alere Triage¿ TOX Drug Screen Control 2 consist of human urine matrix containing parent compounds and/or major metabolites/analogues of drugs of abuse. 510(k)/IDE/PMA number: K060788 Type of Packaging: 5 vials per level in a labeled kit box. Indications for use: The Alere Triage¿ TOX Drug Screen Controls are to be used with the Alere Triage¿ TOX Drug Screen tests and Alere Triage¿ Meters to assist the laboratory in monitoring test performance.
Reason for recall
Alere San Diego, Inc. initiated a voluntary recall of Alere Triage TOX Drug Screen Control 1 (PN 94413, Lot C2681A) because the CODE CHIP that was provided with this lot causes the control to fail when run in the QC Sample Mode using Alere Triage TOX Drug Screeen products. The Alere Triage TOX Drug Screen test devices and patient results are not affected.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0053-2014. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.