Brand Name: MPR Product Name:SINGLE USE TOURNIQUET CUFF 24" 1 LINE — Class II medical device recall Z-0053-2027
Ongoing recall by ASP Global, LLC., reported 2026-10-07, distributed nationwide. Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are la
| Recall number | Z-0053-2027 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | ASP Global, LLC., Austell, GA, United States |
| Recall initiated | 2026-08-24 |
| Classified | 2026-10-01 |
| Reported by FDA | 2026-10-07 |
| Distributed | Nationwide (US) |
| Quantity | 1,740 |
| Reason classes | labeling |
| Model numbers | DT-7024-01Q |
Product
Brand Name: MPR Product Name:SINGLE USE TOURNIQUET CUFF 24" 1 LINE (10) Model/Catalog Number: DT-7024-01Q Software Version: Not Applicable Product Description: Sterile Disposable Tourniquet Cuff is used to be wrapped around a patient's limb and inflated to reduce or totally occlude circulation during surgery. It is non-invasive device. Tourniquet Cuffs are intended for use in surgical procedures that requires the interruption of blood flow and clear visualization of the operative field. Component: No
Reason for recall
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0053-2027. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-10-13. Not affiliated with or endorsed by FDA. Not medical or legal advice.