Data Foundry

Medical device recalls 2016

Vascular Solutions Twin-Pass RX Dual Access Catheter — Class I medical device recall Z-0054-2017

Terminated recall by Vascular Solutions, Inc., reported 2016-10-26, distributed nationwide. Vascular Solutions is recalling Twin-Pass dual access catheters because there is a potential for excess manufa

Recall numberZ-0054-2017
Product typeMedical device
ClassificationClass I
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmVascular Solutions, Inc., Maple Grove, MN, United States
Recall initiated2016-09-16
Classified2016-10-20
Reported by FDA2016-10-26
Terminated2018-05-04
DistributedNationwide (US)
Countries outside the USAT, AU, BE, BR, CA, CH, CY, CZ, DE, EG, ES, FI, FR, GB, GR, HK, HU, IE, IL, IT, NL, NO, PL, SA, SE, TR, TW, ZA
Quantity11,670 units
Lot numbers577762
Model numbers5200, 5210, 5230

Product

Vascular Solutions Twin-Pass RX Dual Access Catheter, model 5210. Sterilized with Ethylene Oxide. The Twin-Pass dual access catheter is a dual lumen catheter designed for use in the coronary and/or peripheral vasculature. The Twin-Pass catheter (Models 5200 and 5230) consists of an over-the-wire (OTW) lumen that runs the length of the catheter and a rapid exchange (RX) delivery lumen on the distal segment. The Twin-Pass catheter has a working length of 135cm and is compatible with various guidewire diameters. Model 5200 also has hydrophilic coating on the distal 18cm of the catheter. The Twin-Pass catheter has white positioning marks located at 95cm (single mark) and 105cm (double marks) from the distal tip, respectively. The Twin-Pass catheter has a marker band located 1mm from the distal tip. Model 5200 contains a second marker band located 11mm from the distal tip to identify the distal end of the OTW lumen.

Reason for recall

Vascular Solutions is recalling Twin-Pass dual access catheters because there is a potential for excess manufacturing material to remain at the tip of the catheter or within the distal portion of the rapid exchange lumen.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0054-2017. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.