NovaSeq 6000Dx including DRAGEN Server — Class II medical device recall Z-0068-2027
Ongoing recall by Illumina, Inc., reported 2026-10-07, distributed nationwide. Retrospectively reported correction for an instrument intended for targeted sequencing of DNA libraries from h
| Recall number | Z-0068-2027 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Illumina, Inc., San Diego, CA, United States |
| Recall initiated | 2024-09-17 |
| Classified | 2026-10-01 |
| Reported by FDA | 2026-10-07 |
| Distributed | Nationwide (US) |
| Countries outside the US | AE, AT, AU, CH, CZ, DE, ES, FI, FR, GB, IT, JP, KR, PL, SE, SG, SK |
| Quantity | 80 |
| UPC / GTIN / UDI-DI | 00816270020637 |
| Serial numbers | 20086130/ACN7003517, ACM4005610, ACM5002946, ACM5003172, ACM5003179, ACM6002015, ACM6002494, ACM6002498, ACM6002503, ACM6002504, ACM9002241, ACM9002243, ACMC007031, ACMC007033, ACMC007048, ACN1003248, ACN1003254, ACN1003255, ACN1003256, ACN1003257, ACN1003258, ACN1003259, ACN1003262, ACN1003268, ACN1003269 and 54 more |
| Model numbers | 20068232, 20086130 |
Product
NovaSeq 6000Dx including DRAGEN Server, Catalog Number: 20068232; Illumina DRAGEN Server for NextSeq 550Dx, Catalog Number: 20086130
Reason for recall
Retrospectively reported correction for an instrument intended for targeted sequencing of DNA libraries from human genomic DNA used with in vitro diagnostic assays has a server with a security vulnerability.