Symbiq One Channel Infuser — Class I medical device recall Z-0069-2013
Terminated recall by Hospira Inc., reported 2012-10-31, distributed nationwide. The Symbiq pump touchscreen may not respond to user selection, may experience a delayed response, or may regis
| Recall number | Z-0069-2013 |
|---|---|
| Product type | Medical device |
| Classification | Class I |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Hospira Inc., Lake Forest, IL, United States |
| Recall initiated | 2012-08-29 |
| Classified | 2012-10-19 |
| Reported by FDA | 2012-10-31 |
| Terminated | 2017-03-15 |
| Distributed | Nationwide (US) |
| Countries outside the US | CA |
| Quantity | 19,614 units |
Product
Symbiq One Channel Infuser; an Rx medical device infusion pump used to administer I.V. fluids; Hospira, Inc., Lake Forest, IL 60045 USA, Made in Costa Rica; List Number 16026-01 Device is a general purpose infusion pump designed to deliver fluids, solutions, medications, agents, nutritionals, electrolytes, blood and blood products for parenteral, enteral, intravenous, intra-arterial, subcutaneous, epidural
Reason for recall
The Symbiq pump touchscreen may not respond to user selection, may experience a delayed response, or may register a different value from the value selected by the user. Failure of the touchscreen to respond to user input could result in a delay in therapy or result in over delivery or under delivery of medication if the user does not confirm the programmed values on the pump's confirmation screen prior to starting the infusion.