Data Foundry

Medical device recalls 2026

VERIFY Dual Species Self-Contained Biological Indicator Model/Catalog… — Class II medical device recall Z-0072-2027

Ongoing recall by Steris Corporation, reported 2026-10-14, distributed nationwide. Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specifi

Recall numberZ-0072-2027
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmSteris Corporation, Mentor, OH, United States
Recall initiated2026-08-31
Classified2026-10-02
Reported by FDA2026-10-14
DistributedNationwide (US)
Quantity1,185
UPC / GTIN / UDI-DI10724995023833
Lot numbers260810, 260810A, 261106, 270116, 270302
Model numbersS3060
Expiration dates2026-08-10, 2026-11-06, 2027-01-16, 2027-03-02

Product

VERIFY Dual Species Self-Contained Biological Indicator Model/Catalog Number: S3060

Reason for recall

Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction