VERIFY Biological Steam Test Pack Model/Catalog Number: S3065 — Class II medical device recall Z-0074-2027
Ongoing recall by Steris Corporation, reported 2026-10-14, distributed nationwide. Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specifi
| Recall number | Z-0074-2027 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Steris Corporation, Mentor, OH, United States |
| Recall initiated | 2026-08-31 |
| Classified | 2026-10-02 |
| Reported by FDA | 2026-10-14 |
| Distributed | Nationwide (US) |
| Quantity | 217 |
| UPC / GTIN / UDI-DI | 10724995220300 |
| Lot numbers | 260810, 260810A, 261106, 270116 |
| Model numbers | S3065 |
| Expiration dates | 2026-08-10, 2026-11-06, 2027-01-16 |
Product
VERIFY Biological Steam Test Pack Model/Catalog Number: S3065
Reason for recall
Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0074-2027. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-10-21. Not affiliated with or endorsed by FDA. Not medical or legal advice.