Data Foundry

Medical device recalls 2017

FlexFlow Venous Cannula Model/Catalog Number 200-200 The FlexFlow… — Class II medical device recall Z-0076-2018

Terminated recall by Sorin Group USA, Inc., reported 2017-11-15, distributed nationwide. A specific lot of the FlexFlowTM Venous Cannula, 200-200, is being recalled because it is sharper or pointier

Recall numberZ-0076-2018
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSorin Group USA, Inc., Arvada, CO, United States
Recall initiated2017-09-28
Classified2017-11-08
Reported by FDA2017-11-15
Terminated2018-10-17
DistributedNationwide (US)
Countries outside the USES, FR
Quantity40 units
Lot numbers181185
Model numbers200-200

Product

FlexFlow Venous Cannula Model/Catalog Number 200-200 The FlexFlow Venous Cannula is a single use sterile device. It is a size 23 French with an outside diameter of 0.30 inch (7.7 mm) and an effective working length of 11.2 inch (28.5 cm). The FlexFlow Venous Cannula is an open lumen polymer tube incorporating wire reinforcement in distal sections. The distal sections of the cannula are perforated with multiple openings to allow increased fluid flow. The clear proximal section is not reinforced to allow clamping the proximal end that terminates in a 3/8 inch (9.5 mm) barbed connector for standard cardiopulmonary bypass tubing. Each cannula is furnished with a mated obturator. Each obturator has a malleable wire. Each component is packaged inside a protective sheath in a single sterile, sealed pouch

Reason for recall

A specific lot of the FlexFlowTM Venous Cannula, 200-200, is being recalled because it is sharper or pointier than expected.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0076-2018. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.