Spectral CT and Spectral CT 7500 RT — Class II medical device recall Z-0078-2027
Ongoing recall by PHILIPS MEDICAL SYSTEMS, reported 2026-10-14, distributed nationwide. Software issues may affect the safety and performance of Spectral CT system(s). The issues may result in image
| Recall number | Z-0078-2027 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | PHILIPS MEDICAL SYSTEMS, Cambridge, MA, United States |
| Recall initiated | 2026-08-21 |
| Classified | 2026-10-02 |
| Reported by FDA | 2026-10-14 |
| Distributed | Nationwide (US) |
| Countries outside the US | AT, AU, BE, BG, CH, CN, CZ, DE, DK, EC, ES, FR, GB, HK, HR, ID, IE, IL, IN, IQ, IT, JO, JP, KR, LT, LV, MX, MY, NI, NL, NO, NZ, PA, PH, PL, PT, RO, SG, SV, TH, TW, VN |
| Quantity | 539 |
| Reason classes | software |
| UPC / GTIN / UDI-DI | 00884838101111, 00884838111103, 00884838117822, 00884838123748, 00884838127876 |
| Serial numbers | 10013, 10015, 10016, 10017, 10018, 10019, 10020, 10021, 10023, 10024, 10025, 10027, 10031, 10032, 10033, 10034, 10035, 10036, 10038, 10039, 10040, 10041, 10042, 10043, 10044 and 475 more |
| Model numbers | 728333, 728340, 728344, 728345 |
Product
Spectral CT and Spectral CT 7500 RT; Software Version Numbers: 5.2.X (U.S.); 5.0.X, 5.2.X, 5.4.X (Outside U.S.); Model Numbers: 1. 728333 (U.S. and Outside U.S.), 2. 728340 (Outside U.S.), 3. 728344 (Outside U.S.), 4. 728345 (Outside U.S.);
Reason for recall
Software issues may affect the safety and performance of Spectral CT system(s). The issues may result in image artifacts, wrong slice thickness, or unintended system motion.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0078-2027. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-10-21. Not affiliated with or endorsed by FDA. Not medical or legal advice.