Presource Packs containing Medtronic Microstream Components:… — Class II medical device recall Z-0079-2027
Ongoing recall by Cardinal Health 200, LLC, reported 2026-10-14. Kits contain Microstream components. Medtronic issued an update to the Instructions for Use of these component
| Recall number | Z-0079-2027 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Cardinal Health 200, LLC, Waukegan, IL, United States |
| Recall initiated | 2026-08-19 |
| Classified | 2026-10-02 |
| Reported by FDA | 2026-10-14 |
| Quantity | 210 kits |
| UPC / GTIN / UDI-DI | 10197106052105, 10197106159828, 10197106704035, 10197106950074, 10198956162259, 10198956177987 |
| Lot numbers | 18464, 18569, 19224, 20866, 23139, 24264, 25795, 27111, 27899, 29023, 29488, 30527, 31037, 31698, 32777, 33326, 33461, 33624, 35231, 35535, 35689, 36554, 36600, 36825, 38327, 38891, 40511, 40711, 41864, 43975, 44933, 45972, 46520, 47547, 47765, 47897, 48088, 48528, 49440, 50521 and 9 more |
| Model numbers | AN23366084, AN23366085, AN2336608G, AN2336608H, AN2336608J, OMSVM4166A |
Product
Presource Packs containing Medtronic Microstream Components: Intubation Kit Catalog Numbers AN23366084 AN23366085 AN2336608H AN2336608J Intubating Bag Kit Catalog Number OMSVM4166A Intubation UH Kit Catalog Number AN2336608G
Reason for recall
Kits contain Microstream components. Medtronic issued an update to the Instructions for Use of these components due reports of difficulty in disconnecting the adapter from the patient's endotracheal tube to perform a procedure. The issue can result in a delay to treatment and unintended extubation. The potential exists for respiratory failure, hypoxia, aspiration, or low oxygen saturation.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0079-2027. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-10-21. Not affiliated with or endorsed by FDA. Not medical or legal advice.