Data Foundry

Medical device recalls 2026

Presource Packs containing Medtronic Microstream Components:… — Class II medical device recall Z-0079-2027

Ongoing recall by Cardinal Health 200, LLC, reported 2026-10-14. Kits contain Microstream components. Medtronic issued an update to the Instructions for Use of these component

Recall numberZ-0079-2027
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmCardinal Health 200, LLC, Waukegan, IL, United States
Recall initiated2026-08-19
Classified2026-10-02
Reported by FDA2026-10-14
Quantity210 kits
UPC / GTIN / UDI-DI10197106052105, 10197106159828, 10197106704035, 10197106950074, 10198956162259, 10198956177987
Lot numbers18464, 18569, 19224, 20866, 23139, 24264, 25795, 27111, 27899, 29023, 29488, 30527, 31037, 31698, 32777, 33326, 33461, 33624, 35231, 35535, 35689, 36554, 36600, 36825, 38327, 38891, 40511, 40711, 41864, 43975, 44933, 45972, 46520, 47547, 47765, 47897, 48088, 48528, 49440, 50521 and 9 more
Model numbersAN23366084, AN23366085, AN2336608G, AN2336608H, AN2336608J, OMSVM4166A

Product

Presource Packs containing Medtronic Microstream Components: Intubation Kit Catalog Numbers AN23366084 AN23366085 AN2336608H AN2336608J Intubating Bag Kit Catalog Number OMSVM4166A Intubation UH Kit Catalog Number AN2336608G

Reason for recall

Kits contain Microstream components. Medtronic issued an update to the Instructions for Use of these components due reports of difficulty in disconnecting the adapter from the patient's endotracheal tube to perform a procedure. The issue can result in a delay to treatment and unintended extubation. The potential exists for respiratory failure, hypoxia, aspiration, or low oxygen saturation.