Positron Emission Tomography and Computed Tomography System — Class II medical device recall Z-0087-2023

Ongoing recall by Shanghai United Imaging Healthcare Co., Ltd., reported 2022-10-19, distributed in CA, MI, TX, WI. The wireless VSM module of a mobile PET/CT system, operating in an environment with strong Wi-Fi signals, may

Recall numberZ-0087-2023
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmShanghai United Imaging Healthcare Co., Ltd., Shanghai, N/A, China
Recall initiated2022-05-30
Classified2022-10-13
Reported by FDA2022-10-19
DistributedCA, MI, TX, WI
Quantity32
UPC / GTIN / UDI-DI06971576832026
Serial numbers06971576832026/230002, 230003, 230004, 230005, 230006, 230007, 230008, 230009, 230010, 230011, 230012, 230013, 230014, 230015, 230016, 230017, 230019, 230020, 230021, 230022, 230023, 230024, 230025, 230026, 230027 and 7 more
Model numbers550

Product

Positron Emission Tomography and Computed Tomography System, Model: uMI 550

Reason for recall

The wireless VSM module of a mobile PET/CT system, operating in an environment with strong Wi-Fi signals, may experience ECG signal and respiratory signal loss due to Wi-Fi interference. ECG and respiratory signal loss during acquisition can result in the failure of ECG and respiratory-gated reconstruction of the PET scan, which may require rescanning of the patient.