Data Foundry

Medical device recalls 2016

VITROS Chemistry Products GLU Slides — Class II medical device recall Z-0089-2017

Terminated recall by Ortho-Clinical Diagnostics, reported 2016-10-26, distributed nationwide. Ortho Clinical Diagnostics confirmed that N-Acetylcysteine (NAC), a drug used for acetaminophen overdose, muco

Recall numberZ-0089-2017
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmOrtho-Clinical Diagnostics, Rochester, NY, United States
Recall initiated2016-08-11
Classified2016-10-17
Reported by FDA2016-10-26
Terminated2018-06-11
DistributedNationwide (US)
Countries outside the USAU, BM, BR, CA, CL, CN, CO, DE, ES, FR, GB, IN, IT, JP, MX, PA, PL, SG, VE
UPC / GTIN / UDI-DI10758750009572

Product

VITROS Chemistry Products GLU Slides, REF/Product Code 170 7801 (300 slides per sales unit), Unique Device Identifier No. 10758750009572, Rx ONLY, IVD --- For in vitro diagnostic use only. VITROS Chemistry Products GLU Slides quantitatively measure glucose (GLU) concentration in serum, plasma, urine, and cerebrospinal fluid using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System

Reason for recall

Ortho Clinical Diagnostics confirmed that N-Acetylcysteine (NAC), a drug used for acetaminophen overdose, mucolytic therapy and is used as nutritional supplement, can interfere with VITROS GLU, LAC, TRIG, and URIC Slides assays. Ortho also tested several compounds following industry standard CLSI EP7-A2 for potential interference on VITROS GLU Slides during validation testing for a proposed change and confirmed Dextran 40 and Glutathione may interfere with current VITROS GLU Slides.