Data Foundry

Medical device recalls 2024

Leica Biosystems Cryostat Model Number CM1950 — Class II medical device recall Z-0089-2025

Ongoing recall by LEICA BIOSYSTEMS NUSSLOCH GMBH, reported 2024-10-23, distributed nationwide. As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays

Recall numberZ-0089-2025
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmLEICA BIOSYSTEMS NUSSLOCH GMBH, Nussloch, Germany
Recall initiated2024-09-04
Classified2024-10-15
Reported by FDA2024-10-23
DistributedNationwide (US)
Quantity30 units
UPC / GTIN / UDI-DI04049188002464, 04049188002488, 04049188002501, 04049188053992
Serial numbers215012008, 216012008, 218012008, 249032008, 287032008, 288032008, 291042008, 314042008, 318042008, 328052008, 335, 356062008, 380, 391072008, 398, 399072008, 407072008, 420, 427082008, 429, 432082008, 435082008, 451082008, 454, 476 and 5 more
Model numbersCM1950

Product

Leica Biosystems Cryostat Model Number CM1950, Material Numbers: a) 14047742467, b) 1491950C1US, c) 1491950C2US, d) 1491950C3US; for freezing and sectioning tissue samples

Reason for recall

As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potentially ignite and therefore can cause serious injuries. The flammability is indicated on the labelling of these sprays. The cryostat labelling, however, did not include any specific warning about usage of flammable freezing sprays prior to 2019.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0089-2025. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.