REF 5100-088 — Class II medical device recall Z-0092-2014
Terminated recall by Stryker Instruments Div. of Stryker Corporation, reported 2013-11-06, distributed nationwide. It was discovered that the silicone potting failed to properly cure after undergoing the prescribed curing pos
| Recall number | Z-0092-2014 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Stryker Instruments Div. of Stryker Corporation, Portage, MI, United States |
| Recall initiated | 2013-10-02 |
| Classified | 2013-10-28 |
| Reported by FDA | 2013-11-06 |
| Terminated | 2014-04-15 |
| Distributed | Nationwide (US) |
| Countries outside the US | AU, CA, CH, DE, ES, FR, GB, IT, JP |
| Quantity | 67 |
| Reason classes | packaging |
| Model numbers | 5100-088 |
Product
REF 5100-088, TPS Micro Driver, For Use with Stryker Total Performance System Caution Federal Law (USA) restricts the device to use by or on the order of a physician Driver, Wire and Bone Drill; The Stryker Total Performance System (TPS) is intended for use in the cutting, drilling, reaming, decorticating and smoothing of bone and other bone related tissues for a variety of applications such as ENT, dental, orthopedic, maxillofacial, spinal, and plastic surgery. The instruments are also used in the placement of screws, wires, pins, and other fixation devices.
Reason for recall
It was discovered that the silicone potting failed to properly cure after undergoing the prescribed curing possess. The potting material is used to insulate, stabilize and provide a moisture barrier to the TPS MicroDriver main Printed Circuit Board Assembly (PCBA). Inadequate cure of the silicone potting could potentially allow it to leak to the outside of the handpiece or cause an electrical ma