Data Foundry

Medical device recalls 2013

REF 5100-088 — Class II medical device recall Z-0092-2014

Terminated recall by Stryker Instruments Div. of Stryker Corporation, reported 2013-11-06, distributed nationwide. It was discovered that the silicone potting failed to properly cure after undergoing the prescribed curing pos

Recall numberZ-0092-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmStryker Instruments Div. of Stryker Corporation, Portage, MI, United States
Recall initiated2013-10-02
Classified2013-10-28
Reported by FDA2013-11-06
Terminated2014-04-15
DistributedNationwide (US)
Countries outside the USAU, CA, CH, DE, ES, FR, GB, IT, JP
Quantity67
Reason classespackaging
Model numbers5100-088

Product

REF 5100-088, TPS Micro Driver, For Use with Stryker Total Performance System Caution Federal Law (USA) restricts the device to use by or on the order of a physician Driver, Wire and Bone Drill; The Stryker Total Performance System (TPS) is intended for use in the cutting, drilling, reaming, decorticating and smoothing of bone and other bone related tissues for a variety of applications such as ENT, dental, orthopedic, maxillofacial, spinal, and plastic surgery. The instruments are also used in the placement of screws, wires, pins, and other fixation devices.

Reason for recall

It was discovered that the silicone potting failed to properly cure after undergoing the prescribed curing possess. The potting material is used to insulate, stabilize and provide a moisture barrier to the TPS MicroDriver main Printed Circuit Board Assembly (PCBA). Inadequate cure of the silicone potting could potentially allow it to leak to the outside of the handpiece or cause an electrical ma

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0092-2014. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.