Data Foundry

Medical device recalls 2016

VITROS Chemistry Products URIC Slides — Class II medical device recall Z-0092-2017

Terminated recall by Ortho-Clinical Diagnostics, reported 2016-10-26, distributed nationwide. Ortho Clinical Diagnostics confirmed that N-Acetylcysteine (NAC), a drug used for acetaminophen overdose, muco

Recall numberZ-0092-2017
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmOrtho-Clinical Diagnostics, Rochester, NY, United States
Recall initiated2016-08-11
Classified2016-10-17
Reported by FDA2016-10-26
Terminated2018-06-11
DistributedNationwide (US)
Countries outside the USAU, BM, BR, CA, CL, CN, CO, DE, ES, FR, GB, IN, IT, JP, MX, PA, PL, SG, VE
UPC / GTIN / UDI-DI10758750000340

Product

VITROS Chemistry Products URIC Slides, REF/Product Code 194 3927 (300 slides per sales unit), Unique Device Identifier No. 10758750000340, Rx ONLY, IVD --- For in vitro diagnostic use only.

Reason for recall

Ortho Clinical Diagnostics confirmed that N-Acetylcysteine (NAC), a drug used for acetaminophen overdose, mucolytic therapy and is used as nutritional supplement, can interfere with VITROS GLU, LAC, TRIG, and URIC Slides assays.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0092-2017. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.