Data Foundry

Medical device recalls 2024

Boston Scientific AVVIGO+ Multi-Modality Guidance System — Class II medical device recall Z-0095-2025

Ongoing recall by Boston Scientific Corporation, reported 2024-10-23, distributed nationwide. Boston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not pr

Recall numberZ-0095-2025
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmBoston Scientific Corporation, Maple Grove, MN, United States
Recall initiated2024-09-05
Classified2024-10-16
Reported by FDA2024-10-23
DistributedNationwide (US)
Quantity9 units
Reason classessoftware
UPC / GTIN / UDI-DI00191506033224
Serial numbers102888346, 102888349, 103233723, 103233724, 104384572, 104399815, 104420529, 104618017, 104618220

Product

Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO+ BRIDGE TO INTEGRATED KIT; the system utilizes an interactive display to provide physicians with physiologic, intravascular and/or intracardiac anatomical information. The system can provide various modalities which includes intravascular ultrasound (IVUS). When used with a peripheral vascular imaging catheter, the system utilizes the IVUS modality for ultrasound examination of intravascular pathology.

Reason for recall

Boston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System while in either Live or Record mode. The Boston Scientific investigation determined that, due to a software anomaly, a 9-grid mark overlay is incorrectly displayed during Live mode when AVVIGO+ is connected to an OptiCross 18 Peripheral Imaging Catheter. Note that the correct overlay display for this catheter is the 15-grid mark overlay, which allows for a reference depth setting of 14.9mm for peripheral catheters. This software anomaly is limited to Live or Record mode. In Review mode, the correct 15-grid mark overlay is displayed, and the vessel image is correctly scaled to the appropriate size.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0095-2025. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.