The GemStar Power Supply is an accessory for the GemStar Infusion Pump — Class I medical device recall Z-0096-2015
Terminated recall by Hospira Inc., reported 2014-11-05, distributed in AK, AL, AR, AZ, CA, CO, CT, DE…. The GemStar Infusion pump may not receive direct current (DC) power from the power supply. In one instance, sm
| Recall number | Z-0096-2015 |
|---|---|
| Product type | Medical device |
| Classification | Class I |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Hospira Inc., Lake Forest, IL, United States |
| Recall initiated | 2014-02-03 |
| Classified | 2014-10-24 |
| Reported by FDA | 2014-11-05 |
| Terminated | 2017-03-07 |
| Distributed | AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY |
| Countries outside the US | AT, BE, BR, CH, DE, ES, FR, GB, GR, NL, SA, SE, TR, ZA |
| Lot numbers | 10115G1, 11121G1, 11122G1, 11124G1, 11128G1, 1123G1, 12140G1, 12142G1, 12143G1, 12144G1, 12148G1, 12178G1, 12195G1, 14176G1, 15185G1, 15187G1, 15196G1, 16207G1, 16208G1, 17214G1, 18217G1, 18218G1, 18219G1, 19225G1, 19226G1, 19229G1, 20233G1, 20234G1, 21241G1, 21242G1, 22251G1, 22255G1, 23259G1, 23262G1, 24270G1, 24271G1, 25276G1, 25280G1, 26285G1, 26289G1 and 6 more |
Product
The GemStar Power Supply is an accessory for the GemStar Infusion Pump. It is a two piece assembled switching power supply. The Power Supply converts alternating current (AC) line input voltage of 120 or 240 V to an output to the device of 3.3 Volts DC.
Reason for recall
The GemStar Infusion pump may not receive direct current (DC) power from the power supply. In one instance, smoke was observed and the GemStar pump was operating on battery power while connected to the 3 volt DC power supply.