Data Foundry

Medical device recalls 2023

Constellation Vision System — Class II medical device recall Z-0099-2024

Ongoing recall by Alcon Research, LLC, reported 2023-10-18, distributed in GA, IL, PR. Ophthalmic microsurgical vision systems have a new Printed Circuit Board (PCB) design that could lead to highe

Recall numberZ-0099-2024
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmAlcon Research, LLC, Irvine, CA, United States
Recall initiated2023-08-18
Classified2023-10-11
Reported by FDA2023-10-18
DistributedGA, IL, PR
Quantity3
Reason classesmicrobial contamination
UPC / GTIN / UDI-DI00380657511501
Serial numbers0802983301X, 1202865201X, 1403028701X
Model numbers8065751150

Product

Constellation Vision System, REF: 8065751150

Reason for recall

Ophthalmic microsurgical vision systems have a new Printed Circuit Board (PCB) design that could lead to higher voltage to a PCB component, that could cause component failure and loss of pressure control for the use of cutters, scissors, and forceps, which could lead to an increased risk of microbial infection, tissue damage and progressive visual impairment.