Data Foundry

Medical device recalls 2025

Model Number U125 — Class I medical device recall Z-0101-2026

Ongoing recall by Boston Scientific Corporation, reported 2025-10-22. Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an a

Recall numberZ-0101-2026
Product typeMedical device
ClassificationClass I
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmBoston Scientific Corporation, Saint Paul, MN, United States
Recall initiated2025-08-20
Classified2025-10-10
Reported by FDA2025-10-22
Quantity3,620 units
Reason classessoftware
UPC / GTIN / UDI-DI00802526559389, 00802526559396, 00802526577109, 00802526578793
Serial numbers100002, 100003, 100004, 100005, 100006, 100007, 100008, 100009, 100010, 100011, 100013, 100015, 100020, 100024, 100026, 100028, 100029, 100033, 100050, 100054, 100057, 100058, 100062, 100063, 100066 and 475 more
Model numbersU125

Product

Model Number U125, VALITUDE CRT-P EL

Reason for recall

Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0101-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.