Data Foundry

Medical device recalls 2022

CENTURION SnagFree Chest Tube insertion Tray — Class II medical device recall Z-0106-2023

Terminated recall by MEDLINE INDUSTRIES, LP - Northfield, reported 2022-10-26, distributed nationwide. One lot of chest tube insertion trays, containing a silk suture component that is restricted to 1 cycle of ste

Recall numberZ-0106-2023
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmMEDLINE INDUSTRIES, LP - Northfield, Northfield, IL, United States
Recall initiated2022-09-07
Classified2022-10-20
Reported by FDA2022-10-26
Terminated2026-03-04
DistributedNationwide (US)
Quantity18 kits
UPC / GTIN / UDI-DI20653160222979
Lot numbers2022050690

Product

CENTURION SnagFree Chest Tube insertion Tray, Reorder CHT545

Reason for recall

One lot of chest tube insertion trays, containing a silk suture component that is restricted to 1 cycle of sterilization, was processed though 2 cycles.