Data Foundry

Medical device recalls 2013

Siemens RAPIDLab¿ 1260 system blood gas analyzer Product Usage:… — Class II medical device recall Z-0107-2014

Terminated recall by Siemens Healthcare Diagnostics Inc, reported 2013-11-06, distributed nationwide. RAPIDPoint¿ 400 Series, RAPIDPoint¿ 500, or RAPIDLab¿ 1200 Series blood gas analyzer, users can inadvertently

Recall numberZ-0107-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSiemens Healthcare Diagnostics Inc, Norwood, MA, United States
Recall initiated2012-09-17
Classified2013-10-29
Reported by FDA2013-11-06
Terminated2015-04-01
DistributedNationwide (US)
Countries outside the USAE, AR, AT, AU, BD, BE, BG, BH, BR, CH, CL, CN, CO, CZ, DE, DK, DZ, EE, EG, ES, FI, FR, GB, GR, HK, HR, HU, ID, IE, IL, IN, IQ, IS, IT, JP, KE, KR, KW, LT, LU, LV, MY, NL, NO, NZ, PH, PL, PT, QA, RO, RS, SA, SE, SG, SI, SK, TH, TR, TW, VE, VN, ZA
Quantity140 units

Product

Siemens RAPIDLab¿ 1260 system blood gas analyzer Product Usage: RAPIDLab 1200 Systems: The Rapidlab 1200 systems are intended for in vitro diagnostic use by healthcare professionals in the quantitative testing of human whole blood. The systems can determine the following parameters: RAPIDLab 1240 pH, pC02, p02 RAPIDLab 1245 pH, pC02, p02, tHb, nBili, F02Hb, FCOHb, FMetHb, FHHb RAPIDLab 1260 pH, pC02, p02, Na+, K+, Ca++, Cl-, glucose, lactate RAPIDLab 1265 pH, pC02, p02, Na+, K+, Ca++, Cl-, glucose, lactate, tHb, nBili, F02Hb, FCOHb, FMetHb, FHHb

Reason for recall

RAPIDPoint¿ 400 Series, RAPIDPoint¿ 500, or RAPIDLab¿ 1200 Series blood gas analyzer, users can inadvertently enter a wrong patient ID on the demographics screen, which ultimately could lead to a misidentification of the sample