SEDIPLAST — Class II medical device recall Z-0107-2027
Ongoing recall by POLYMEDCO, INC. / POLYMEDCO CANCER DIAGNOSTIC PRODUCTS, LLC, reported 2026-10-14, distributed nationwide. During use of an affected tube, the operator may encounter increased resistance when inserting the pipette. Th
| Recall number | Z-0107-2027 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | POLYMEDCO, INC. / POLYMEDCO CANCER DIAGNOSTIC PRODUCTS, LLC, Cortlandt Manor, NY, United States |
| Recall initiated | 2026-08-25 |
| Classified | 2026-10-07 |
| Reported by FDA | 2026-10-14 |
| Distributed | Nationwide (US) |
| Countries outside the US | GB |
| Quantity | 448,400 |
| Reason classes | device malfunction |
| Lot numbers | E1150A |
| Model numbers | S-100, S-1000 |
| Expiration dates | 2027-06-30 |
Product
SEDIPLAST; Model/Catalog Numbers: S-100 and S-1000;
Reason for recall
During use of an affected tube, the operator may encounter increased resistance when inserting the pipette. The resulting pressure buildup may cause the tube cap to detach or be forcibly ejected, potentially resulting in the release of blood and exposure of the operator or surrounding work area to a biological hazard (blood).
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0107-2027. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-10-21. Not affiliated with or endorsed by FDA. Not medical or legal advice.