Data Foundry

Medical device recalls 2026

SEDIPLAST — Class II medical device recall Z-0107-2027

Ongoing recall by POLYMEDCO, INC. / POLYMEDCO CANCER DIAGNOSTIC PRODUCTS, LLC, reported 2026-10-14, distributed nationwide. During use of an affected tube, the operator may encounter increased resistance when inserting the pipette. Th

Recall numberZ-0107-2027
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmPOLYMEDCO, INC. / POLYMEDCO CANCER DIAGNOSTIC PRODUCTS, LLC, Cortlandt Manor, NY, United States
Recall initiated2026-08-25
Classified2026-10-07
Reported by FDA2026-10-14
DistributedNationwide (US)
Countries outside the USGB
Quantity448,400
Reason classesdevice malfunction
Lot numbersE1150A
Model numbersS-100, S-1000
Expiration dates2027-06-30

Product

SEDIPLAST; Model/Catalog Numbers: S-100 and S-1000;

Reason for recall

During use of an affected tube, the operator may encounter increased resistance when inserting the pipette. The resulting pressure buildup may cause the tube cap to detach or be forcibly ejected, potentially resulting in the release of blood and exposure of the operator or surrounding work area to a biological hazard (blood).