St Jude Medical — Class II medical device recall Z-0108-2015
Terminated recall by St Jude Medical, reported 2014-10-29, distributed nationwide. St. Jude Medical is conducting a voluntary recall of SMJ Batch # 4641580 of the Fast-Cath Trio" Hemostasis Int
| Recall number | Z-0108-2015 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | St Jude Medical, Saint Paul, MN, United States |
| Recall initiated | 2014-10-14 |
| Classified | 2014-10-21 |
| Reported by FDA | 2014-10-29 |
| Terminated | 2015-03-25 |
| Distributed | Nationwide (US) |
| Quantity | 220 |
| Lot numbers | 4641580 |
| Model numbers | 406308 |
Product
St Jude Medical, 10F Fast-Cath Trio" Hemostasis Introducer, REF # 406308, Sterile EO. The Fast-Cath Trio" Hemostasis Introducer consists of the following components: Port hemostasis adapter (3), Sheath (10F Cath-Lock"), dilator (10F) and 0.038 guidewire. See product labeling for diagram of contents.
Reason for recall
St. Jude Medical is conducting a voluntary recall of SMJ Batch # 4641580 of the Fast-Cath Trio" Hemostasis Introducer, Catalog number 406308. The inner diameter of the 10F dilator is too small, such that the guidewire cannot advance through the device which may result in a procedural delay while another dilator of the desired size is obtained.