Alphenix INFX-8000C/B5 — Class II medical device recall Z-0108-2027
Ongoing recall by Canon Medical System, USA, INC., reported 2026-10-14, distributed nationwide. Loosening of the bolts securing the longitudinal movement mechanism may cause the tabletop to shift, resulting
| Recall number | Z-0108-2027 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Canon Medical System, USA, INC., Tustin, CA, United States |
| Recall initiated | 2026-08-26 |
| Classified | 2026-10-07 |
| Reported by FDA | 2026-10-14 |
| Distributed | Nationwide (US) |
| Quantity | 1 units |
| UPC / GTIN / UDI-DI | 04987670100192 |
| Serial numbers | B5F2532011 |
Product
Alphenix INFX-8000C/B5
Reason for recall
Loosening of the bolts securing the longitudinal movement mechanism may cause the tabletop to shift, resulting in [positional/imaging/patient] misalignment and a longitudinal motion error