Babylog VN500 Ventilator with Installed CO2 Measurement… — Class II medical device recall Z-0109-2022
Ongoing recall by Draeger Medical, Inc., reported 2021-10-20, distributed nationwide. Software 2.51 and Lower with Installed CO2 Measurement Option, may result in Restart of the Ventilator, brief
| Recall number | Z-0109-2022 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Draeger Medical, Inc., Telford, PA, United States |
| Recall initiated | 2021-09-10 |
| Classified | 2021-10-13 |
| Reported by FDA | 2021-10-20 |
| Distributed | Nationwide (US) |
| Quantity | 2,169 units |
| Reason classes | software |
| Model numbers | 8417400 |
Product
Babylog VN500 Ventilator with Installed CO2 Measurement Option-intended for the ventilation of neonatal patients from 0.4 kg (0.88 lbs) up to 10 kg (22 lbs) and pediatric patients from 5 kg (11 lbs) up to 20 kg (44 lbs) bodyweight. Catalog Number: 8417400
Reason for recall
Software 2.51 and Lower with Installed CO2 Measurement Option, may result in Restart of the Ventilator, brief cessation of ventilation can occur and loss of PEEP for approx. 8 seconds