Data Foundry

Medical device recalls 2021

Babylog VN500 Ventilator with Installed CO2 Measurement… — Class II medical device recall Z-0109-2022

Ongoing recall by Draeger Medical, Inc., reported 2021-10-20, distributed nationwide. Software 2.51 and Lower with Installed CO2 Measurement Option, may result in Restart of the Ventilator, brief

Recall numberZ-0109-2022
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmDraeger Medical, Inc., Telford, PA, United States
Recall initiated2021-09-10
Classified2021-10-13
Reported by FDA2021-10-20
DistributedNationwide (US)
Quantity2,169 units
Reason classessoftware
Model numbers8417400

Product

Babylog VN500 Ventilator with Installed CO2 Measurement Option-intended for the ventilation of neonatal patients from 0.4 kg (0.88 lbs) up to 10 kg (22 lbs) and pediatric patients from 5 kg (11 lbs) up to 20 kg (44 lbs) bodyweight. Catalog Number: 8417400

Reason for recall

Software 2.51 and Lower with Installed CO2 Measurement Option, may result in Restart of the Ventilator, brief cessation of ventilation can occur and loss of PEEP for approx. 8 seconds

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0109-2022. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.