Data Foundry

Medical device recalls 2012

EVICEL Fibrin Sealant — Class II medical device recall Z-0111-2013

Terminated recall by Ethicon, Inc., reported 2012-10-31, distributed nationwide. Ethicon and Omrix Biopharmaceuticals have received reports of air or gas embolism occuring in association with

Recall numberZ-0111-2013
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmEthicon, Inc., Somerville, NJ, United States
Recall initiated2012-07-02
Classified2012-10-22
Reported by FDA2012-10-31
Terminated2015-09-15
DistributedNationwide (US)
Countries outside the USBR, CA, IL, MX, SG

Product

EVICEL Fibrin Sealant (Human) Rx Only. Manufacturer: Omrix Biopharmaceuticals Ltd. MDA blood bank, Sheba Hospital, Ramat-Gan, POB 888, Kiryat Ono 55000 Israel U.S. License No. 1603. Distributed by: Ethicon, Inc. P.O. Box 151, Somerville, NJ 08876-0151 USA. The Pressure Regulator System is positioned on the floor or on an IV pole and accepts medical gas (compressed air, nitrogen, CO2) from the central hospital system or from medical gas cylinders.

Reason for recall

Ethicon and Omrix Biopharmaceuticals have received reports of air or gas embolism occuring in association with the use of spray devices employing a pressure regulator to administer EVICEL; these events appear to be related to the use of the spray device at higher-than-recommended pressures and/or in close proximity to the tissue surface. The field correction initiated is to provide the recommend