Data Foundry

Medical device recalls 2020

Atellica IM HER-2/neu — Class II medical device recall Z-0115-2021

Terminated recall by Siemens Healthcare Diagnostics, Inc, reported 2020-10-28, distributed nationwide. Siemens confirmed that the Upper Limit of Normal (ULN) as claimed in the Instructions for Use (IFU) was no lon

Recall numberZ-0115-2021
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSiemens Healthcare Diagnostics, Inc, East Walpole, MA, United States
Recall initiated2020-08-25
Classified2020-10-16
Reported by FDA2020-10-28
Terminated2023-03-10
DistributedNationwide (US)
UPC / GTIN / UDI-DI00630414471228, 00630414471235, 00630414599151, 00630414599168
Lot numbers22124A49, 22695A49, 24043138, 32477138, 32478139, 43157138, 64661A48, 81211A48, 81653138, 91743A48, 98034139, 99728138
Model numbers10308993, 10308994, 10995591, 10995592
Expiration dates2020-11-12, 2020-12-11, 2021-01-30

Product

Atellica IM HER-2/neu (H2n) assay, 50 Test Kit, Catalog No. SMN 10995591, UDI Codes: (01)00630414599151(10)98034139(17)20210130 and (01)00630414599151(10)32478139(17)20210130 Atellica IM HER-2/neu (H2n) calibrator, 2 Pack, Catalog No. SMN 10995592, UDI Code (01)00630414599168(10)22695A49(17)20201211 ADVIA Centaur HER-2/neu (H2n) assay, 50 Test Kit, Catalog No. SMN 10308994, UDI codes : (01)00630414471235(10)43157138(17)20210130 (01)00630414471235(10)81653138(17)20210130 (01)00630414471235(10)99728138(17)20210130 (01)00630414471235(10)24043138(17)20210130 (01)00630414471235(10)32477138(17)20210130 Advia Centaur Systems HER-2/neu (H2n) calibrator, 2 Pack, Catalog No. SMN 10308993, UDI Codes: (01)00630414471228(10)64661A48(17)20201112 (01)00630414471228(10)81211A48(17)20201112 (01)00630414471228(10)91743A48(17)20201112 (01)00630414471228(10)22124A49(17)20201211 Product Usage: This product is intended to be used for in vitro diagnostic use in calibrating the ADVIA Centaur systems HER-2/neu assay.

Reason for recall

Siemens confirmed that the Upper Limit of Normal (ULN) as claimed in the Instructions for Use (IFU) was no longer achieved. This can potentially result in false positive or false negative results.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0115-2021. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.