UniCel¿ DxC 600/800 with Software v5 — Class II medical device recall Z-0123-2013
Terminated recall by Beckman Coulter Inc., reported 2012-10-31, distributed in OH, TX. Beckman Coulter has confirmed that the Shuttle to Barcode Reader and Cap Piercer alignment procedure cannot be
| Recall number | Z-0123-2013 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Beckman Coulter Inc., Brea, CA, United States |
| Recall initiated | 2012-09-03 |
| Classified | 2012-10-24 |
| Reported by FDA | 2012-10-31 |
| Terminated | 2014-02-14 |
| Distributed | OH, TX |
| Countries outside the US | CN, TR |
| Quantity | 7 units |
| Reason classes | software |
| Serial numbers | 2329, 5477, 5479 |
| Model numbers | B12218 |
Product
UniCel¿ DxC 600/800 with Software v5.0.11, Part Number B12218. The UniCel¿ DxC Synchron¿ Clinical Systems are fully automated, computercontrolled clinical chemistry analyzers designed for the in vitro determination of a variety of general chemistries, therapeutic drugs, and other chemistries. Analysis can be performed on serum, plasma, urine, or cerebrospinal fluid (CSF) and whole blood (sample type is chemistry dependent).
Reason for recall
Beckman Coulter has confirmed that the Shuttle to Barcode Reader and Cap Piercer alignment procedure cannot be performed on certain DxC configurations with v5.0.11 software installed. If the Shuttle to Barcode Reader becomes misaligned, the Operator will not be able to perform this alignment procedure on the DxC systems listed above. Software v5.0 .11 does not allow access to the Shuttle to Bar