VOCSN+Pro — Class I medical device recall Z-0123-2027
Ongoing recall by Ventec Life Systems, Inc., reported 2026-10-28, distributed nationwide. Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resultin
| Recall number | Z-0123-2027 |
|---|---|
| Product type | Medical device |
| Classification | Class I |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Ventec Life Systems, Inc., Bothell, WA, United States |
| Recall initiated | 2026-09-17 |
| Classified | 2026-10-20 |
| Reported by FDA | 2026-10-28 |
| Distributed | Nationwide (US) |
| Quantity | 325 units |
| UPC / GTIN / UDI-DI | 00855573007204, 00855573007761 |
| Serial numbers | 00855573007204, 00855573007761 |
| Model numbers | PRT-00490-100 |
Product
VOCSN+Pro (Japan), Model Number: PRT-00490-100, Package PN:PRT-00853-100. used for ventilatory support.
Reason for recall
Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0123-2027. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-11-03. Not affiliated with or endorsed by FDA. Not medical or legal advice.