Ti-Magna Fx Washer Ti-Versa-FX¿II Femoral Fixation System S/C Plates… — Class II medical device recall Z-0129-2014
Terminated recall by Zimmer, Inc., reported 2013-11-13. Zimmer recently conducted a review of historic packaging validations completed. Based on this review, it was d
| Recall number | Z-0129-2014 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Zimmer, Inc., Warsaw, IN, United States |
| Recall initiated | 2013-08-28 |
| Classified | 2013-11-01 |
| Reported by FDA | 2013-11-13 |
| Terminated | 2015-04-30 |
| Countries outside the US | JP |
| Quantity | 2,632 devices |
| Reason classes | sterility |
Product
Ti-Magna Fx Washer Ti-Versa-FX¿II Femoral Fixation System S/C Plates The compression tube/plates of this trauma system are used with Ti-Versa-Fx¿ II Femoral Fixation System lag screws for the internal fixation of proximal femoral and supracondylar fractures. The Ti-Magna Fx System is an osteosynthesis item for orthopedic use in open internal fixation of fractures of the femoral neck, condyle, or tibia, etc. The washer prevents the screw head from being buried in the bone.
Reason for recall
Zimmer recently conducted a review of historic packaging validations completed. Based on this review, it was determined that a subset of product packaged at a single Zimmer facility required testing to ensure that previously packaged product from this facility met the established requirements for sterile barrier integrity. The number of samples required for packaging testing, to provide statisti
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0129-2014. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.