Edwards Lifesciences Fem-Flex Femoral Arterial Cannula with Duraflo… — Class II medical device recall Z-0134-2014
Terminated recall by Edwards Lifesciences, LLC, reported 2013-11-13, distributed nationwide. Edwards Lifesciences is recalling two lots of Duraflo Cannulas due to reported complaints involving cracks in
| Recall number | Z-0134-2014 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Edwards Lifesciences, LLC, Draper, UT, United States |
| Recall initiated | 2013-08-26 |
| Classified | 2013-11-01 |
| Reported by FDA | 2013-11-13 |
| Terminated | 2017-02-22 |
| Distributed | Nationwide (US) |
| Quantity | 80 units |
| Reason classes | packaging |
| Lot numbers | 59388992 |
| Expiration dates | 2016-01-01 |
Product
Edwards Lifesciences Fem-Flex Femoral Arterial Cannula with Duraflo Coating, DIITFA02225, Sterile, RX only For Single Use Only, Product Usage: Edwards Lifesciences femoral access cannulae are intended for use in situations which require rapid femoral venous and arterial access for short-term (< 6 hours) cardiopulmonary bypass. Vessel access (venous or arterial) is left to the discretion of the physician. Femoral access cannulae may be used in pediatric populations or adult populations based on flow rate requirements and individual patient anatomy. Please consult labeling to determine pressure drop related to flow rates. Extracorporeal circuit components with a Duraflo coating are intended for use in cardiopulmonary surgery when a heparin coated blood path is desired.
Reason for recall
Edwards Lifesciences is recalling two lots of Duraflo Cannulas due to reported complaints involving cracks in the t-connector of the devices which were confirmed during product evaluation.