Soltive Pro SuperPulsed Laser Systems-Intended for incision — Class II medical device recall Z-0135-2025
Ongoing recall by Olympus Corporation of the Americas, reported 2024-10-30, distributed nationwide. The term Bladder Stone was incorrectly translated in both Spanish and Portuguese to Kidney Stone(Clculo renal)
| Recall number | Z-0135-2025 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Olympus Corporation of the Americas, Center Valley, PA, United States |
| Recall initiated | 2024-08-13 |
| Classified | 2024-10-18 |
| Reported by FDA | 2024-10-30 |
| Distributed | Nationwide (US) |
| Quantity | 159 units |
| Reason classes | foreign material |
| UPC / GTIN / UDI-DI | 00821925044135 |
Product
Soltive Pro SuperPulsed Laser Systems-Intended for incision, excision, resection, ablation, coagulation, hemostasis, and vaporization of soft tissue Model/Catalog Number: TFL-SLS
Reason for recall
The term Bladder Stone was incorrectly translated in both Spanish and Portuguese to Kidney Stone(Clculo renal) on the systems Graphical User Interface (GUI), a potential to deliver an incorrect or unintended amount of energy to the patients anatomy.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0135-2025. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.