Data Foundry

Medical device recalls 2024

GE Healthcare Giraffe Omnibed Carestation CS1 — Class I medical device recall Z-0140-2025

Completed recall by DATEX--OHMEDA, INC., reported 2024-11-06. Prior to current established manufacturing practices, higher levels of airborne formaldehyde exposure from new

Recall numberZ-0140-2025
Product typeMedical device
ClassificationClass I
StatusCompleted
Voluntary or mandatedVoluntary
Recalling firmDATEX--OHMEDA, INC., Wauwatosa, WI, United States
Recall initiated2023-06-06
Classified2024-10-31
Reported by FDA2024-11-06
Quantity769 units
UPC / GTIN / UDI-DI00008406821168, 00840682116862

Product

GE Healthcare Giraffe Omnibed Carestation CS1, combination infant incubator and infant warmer. Model Numbers: 1) 2082844-001-01040418; 2) 2082844-001-01046784; 3) 2082844-001-01052324; 4) 2082844-001-01052572; 5) 2082844-001-01055337; 6) 2082844-001-01065693; 7) 2082844-001-01069008; 8) 2082844-001-01073750; 9) 2082844-001-01086553; 10) 2082844-001-01089393; 11) 2082844-001-01090414; 12) 2082844-001-01092632; 13) 2082844-001-01111482; 14) 2082844-001-01115741; 15) 2082844-001-01122391; 16) 2082844-001-01128579; 17) 2082844-001-01128725; 18) 2082844-001-01131573; 19) 2082844-001-01140642; 20) 2082844-001-01165870; 21) 2082844-001-100397; 22) 2082844-001-100676; 23) 2082844-001-100740; 24) 2082844-001-100743; 25) 2082844-001-100744; 26) 2082844-001-101075; 27) 2082844-001-101149; 28) 2082844-001-101199; 29) 2082844-001-101202; 30) 2082844-001-101203; 31) 2082844-001-101356; 32) 2082844-001-101366; 33) 2082844-001-101656; 34) 2082844-001-102136; 35) 2082844-001-102589; 36) 2082844-001-102858; 37) 2082844-001-526493; 38) 2082844-001-533346; 39) 2082844-001-570904; 40) 2082844-001-615985; 41) 2082844-001-616544; 42) 2082844-001-674814; 43) 2082844-001-701655; 44) 2082844-001-704407; 45) 2082844-001-709042; 46) 2082844-001-857748; 47) 2082844-001-888278; 48) 2082844-001-941288; 49) 2082844-001-965539; 50) 2082844-001-978713; 51) 2082844-001-982099; 52) 2082844-001-982310; 53) 2082844-001-982985; 54) 2082844-001-982988; 55) 2082844-001-983509; 56) 2082844-001-983643; 57) 2082844-001-983679; 58) 2082844-001-984096; 59) 2082844-001-984948; 60) 2082844-001-987480; 61) 2082844-001-987855; 62) 2082844-001-993967; 63) 2082844-001-994441; 64) 2082844-001-995782; 65) 2082844-001-998377; Radiant infant warmer

Reason for recall

Prior to current established manufacturing practices, higher levels of airborne formaldehyde exposure from newly manufactured incubators could be seen during clinical use that would rapidly decrease over a week. Affected customers were notified in June 2023 to run incubators for a week to potentially off gas any residual airborne formaldehyde. The general scientific consensus is that sensory irritation of the eyes is the most sensitive endpoint with respect to exposure to airborne formaldehyde and is the adverse effect clinically experienced at the lowest concentration.