Hillrom Centrella Hospital Bed with WatchCare Incontinence Management… — Class I medical device recall Z-0141-2023
Ongoing recall by Baxter Healthcare Corporation, reported 2022-11-09. Radio frequency emission from a functioning WatchCare device may potentially impact other devices (including,
| Recall number | Z-0141-2023 |
|---|---|
| Product type | Medical device |
| Classification | Class I |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Baxter Healthcare Corporation, Deerfield, IL, United States |
| Recall initiated | 2022-09-30 |
| Classified | 2022-10-31 |
| Reported by FDA | 2022-11-09 |
| Quantity | 6,407 units |
| UPC / GTIN / UDI-DI | 00887761985162 |
| Model numbers | P7900B100193, P7900B100195, P7900B100230, P7900B100236, P7900B100239, P7900B100244, P7900B100256, P7900B100257, P7900B100259, P7900B100264, P7900B100268, P7900B100269, P7900B100292, P7900B100296, P7900B100302, P7900B100303, P7900B100304, P7900B100305, P7900B100312, P7900B100316, P7900B100318, P7900B100335, P7900B100341, P7900B100342, P7900B100354 and 49 more |
Product
Hillrom Centrella Hospital Bed with WatchCare Incontinence Management System, Product code/Part numbers: P7900B100193, P7900B100195, P7900B100230, P7900B100236, P7900B100239, P7900B100244, P7900B100256, P7900B100257, P7900B100259, P7900B100264, P7900B100268, P7900B100269, P7900B100292, P7900B100296, P7900B100302, P7900B100303, P7900B100304, P7900B100305, P7900B100312, P7900B100316, P7900B100318, P7900B100335, P7900B100341, P7900B100342, P7900B100354, P7900B100361, P7900B100368, P7900B100372, P7900B100381, P7900B100382, P7900B100384, P7900B100385, P7900B100387, P7900B100388, P7900B100392, P7900B100395, P7900B100396, P7900B100407, P7900B100408, P7900B100438, P7900B100441, P7900B1PL270, P7900B1PLT71, P7900B1SPL05, P7900B200193, P7900B200195, P7900B200230, P7900B200236, P7900B200264, P7900B200268, P7900B200292, P7900B200302, P7900B200318, P7900B200342, P7900B200361, P7900B200372, P7900B200381, P7900B200382, P7900B200392, P7900B200395, P7900B200396, P7900B200408, P7900B200439, P7900B200441, P7900B200443, P7900B200445, P7900B200452, P7900B200453, P7900B200464, P7900B200485, P7900B2SPL12, P7900B300441, P7900B300453, P7900B300518
Reason for recall
Radio frequency emission from a functioning WatchCare device may potentially impact other devices (including, but not limited to telemetry devices, bladder scanner, fetal monitor/doppler, infusion pumps, and insulin pumps).