Data Foundry

Medical device recalls 2022

Hillrom Centrella Hospital Bed with WatchCare Incontinence Management… — Class I medical device recall Z-0141-2023

Ongoing recall by Baxter Healthcare Corporation, reported 2022-11-09. Radio frequency emission from a functioning WatchCare device may potentially impact other devices (including,

Recall numberZ-0141-2023
Product typeMedical device
ClassificationClass I
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmBaxter Healthcare Corporation, Deerfield, IL, United States
Recall initiated2022-09-30
Classified2022-10-31
Reported by FDA2022-11-09
Quantity6,407 units
UPC / GTIN / UDI-DI00887761985162
Model numbersP7900B100193, P7900B100195, P7900B100230, P7900B100236, P7900B100239, P7900B100244, P7900B100256, P7900B100257, P7900B100259, P7900B100264, P7900B100268, P7900B100269, P7900B100292, P7900B100296, P7900B100302, P7900B100303, P7900B100304, P7900B100305, P7900B100312, P7900B100316, P7900B100318, P7900B100335, P7900B100341, P7900B100342, P7900B100354 and 49 more

Product

Hillrom Centrella Hospital Bed with WatchCare Incontinence Management System, Product code/Part numbers: P7900B100193, P7900B100195, P7900B100230, P7900B100236, P7900B100239, P7900B100244, P7900B100256, P7900B100257, P7900B100259, P7900B100264, P7900B100268, P7900B100269, P7900B100292, P7900B100296, P7900B100302, P7900B100303, P7900B100304, P7900B100305, P7900B100312, P7900B100316, P7900B100318, P7900B100335, P7900B100341, P7900B100342, P7900B100354, P7900B100361, P7900B100368, P7900B100372, P7900B100381, P7900B100382, P7900B100384, P7900B100385, P7900B100387, P7900B100388, P7900B100392, P7900B100395, P7900B100396, P7900B100407, P7900B100408, P7900B100438, P7900B100441, P7900B1PL270, P7900B1PLT71, P7900B1SPL05, P7900B200193, P7900B200195, P7900B200230, P7900B200236, P7900B200264, P7900B200268, P7900B200292, P7900B200302, P7900B200318, P7900B200342, P7900B200361, P7900B200372, P7900B200381, P7900B200382, P7900B200392, P7900B200395, P7900B200396, P7900B200408, P7900B200439, P7900B200441, P7900B200443, P7900B200445, P7900B200452, P7900B200453, P7900B200464, P7900B200485, P7900B2SPL12, P7900B300441, P7900B300453, P7900B300518

Reason for recall

Radio frequency emission from a functioning WatchCare device may potentially impact other devices (including, but not limited to telemetry devices, bladder scanner, fetal monitor/doppler, infusion pumps, and insulin pumps).

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0141-2023. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.