Brand Name: IOMAX Product Name: IOMAX Cortical Module Model/Catalog… — Class II medical device recall Z-0141-2027
Ongoing recall by Cadwell Industries Inc, reported 2026-10-21, distributed nationwide. Stimulator device contains a defective printed circuit board assembly (PCBA) that may cause the device to unin
| Recall number | Z-0141-2027 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Cadwell Industries Inc, Kennewick, WA, United States |
| Recall initiated | 2026-08-18 |
| Classified | 2026-10-09 |
| Reported by FDA | 2026-10-21 |
| Distributed | Nationwide (US) |
| Countries outside the US | BE, BR, CH, CO, ES, IL, IT, MX, MY, OM, SA, TH, TR, VN, ZA |
| Quantity | 23 |
| UPC / GTIN / UDI-DI | 00840067100653 |
| Model numbers | 190296-200 |
Product
Brand Name: IOMAX Product Name: IOMAX Cortical Module Model/Catalog Number: 190296-200 UDI-DI code: 00840067100653 Product Description: The IOMAX Cortical Module contains a 16-channel amplifier, a low current electrical stimulator, a TCS stimulator, a visual stimulator and an auditory stimulator. The low current stimulator has one (1) constant current output pair with a current range of 0.01-20 mA. The TCS stimulator has nine (9) independent constant current/constant voltage outputs, with current range of 50-1500 mA and voltage range of 50-1000 V. Component: Is the cortical component of the IOMAX system
Reason for recall
Stimulator device contains a defective printed circuit board assembly (PCBA) that may cause the device to unintentionally deliver electrical stimulation to the patient, or allow an unintended electrical current to reach the patient.